Prepare for excision of skin tumor, thoroughly clean and dry the skin.
-
Adjust the attachment plates to fit the skin’s curvature by bending the plate.
-
Peel the cover of the adhesive tape underneath the plate.
-
Firmly attach the plate 1.5-2 cm away from the margin of excision.
-
Prepare the medical adhesive tape and apply the tape to secure the attachment plate to the skin.
-
Fix the medical adhesive tape to the skin and peel the covering film.
-
Plan the placement of the contra-lateral attachment plate to align with the vector of pull.
-
Repeat all stages for the second plate.
-
Gently pull the lock/release mechanism and insert the approximation strap.
-
Keep the plate attached to the skin at all times and pull the strap all the way out.
-
Secure the plate to the skin by pressing with the little finger and insert the strap through the lock/release mechanism and pull towards the remote plate.
-
Keep both plates attached to the skin and approximate the margins of excision by pushing both plates towards each other.
-
In case of excessive pull and collapse of the LR/M and the approximation strap wings, re-insert the strap and pull gradually in order to maintain functionality of the system and avoid repeated failure of the approximation strap wings.
-
Lock the strap by pressing down the lock/release mechanism.
-
To release the strap lift up the knob and take the strap out from the other side by pushing it out.
-
Keep the plate down and pull the strap all the way out.
|
|
|
|
-
Prepare for excision of skin tumor, thoroughly clean and dry the skin.
-
Crack the liquid adhesive ampoule and apply to the skin.
-
Allow the liquid adhesive to dry for 20-30 seconds.
-
Adjust the attachment plates to fit the skin’s curvature by bending the plate.
-
Peel the cover of the adhesive tape underneath the plate.
-
Firmly attach the plate 1.5-2 cm away from the margin of excision.
-
Prepare the medical adhesive tape and apply the tape to secure the attachment plate to the skin.
-
Fix the medical adhesive tape to the skin and peel the covering film.
-
Plan the placement of the contra-lateral attachment plate to align with the vector of pull.
-
Repeat all stages for the second plate.
-
Gently pull the lock/release mechanism and insert the approximation strap.
-
Keep the plate attached to the skin at all times and pull the strap all the way out.
-
Secure the plate to the skin by pressing with the little finger and insert the strap through the lock/release mechanism and pull towards the remote plate.
-
Keep both plates attached to the skin and approximate the margins of excision by pushing both plates towards each other.
-
Lock the strap by pressing down the lock/release mechanism.
-
To release the strap lift up the knob and take the strap out from the other side by pushing it out.
-
Keep the plate down and pull the strap all the way out.
|
|
|
|
-
Clean and dry the skin surface prior to application of the attachment plates.
-
Peel off the adhesive tape on the back of the attachment plate.
-
Apply the self-adhesive attachment plate to the wound, 1.5-2 cm from the wound edge.
-
Use staples to further secure the attachment plates to the skin. Please note the size of staples 6.5 mm x 4.7 mm to be compatible with TopClosure®.
-
Apply staples alternating from proximal to distal.
-
Apply the 2nd attachment plate to the opposite side of the wound, 1.5-2mm from the wound's edge, in line with the vector of pull.
-
Secure the attachment plate to the skin using staples.
-
Choose the tension sutures.
-
Apply tension sutures through the designated oval holes in the attachment plate, grip the deep soft tissue layers on both sides of the wound and add a sleeve to the suture to protect the outer layer.
-
Apply repeated Stress-Relaxation to the skin with tension sutures, pull for 30 seconds and relax for 1 minute.
-
Ty the tension sutures in stages.
-
Apply staples to adjust skin approximation.
-
Add the approximation straps for reduction of wound tension for final wound closure.
-
On open wounds sterile gauze must be placed under the approximation strap and if closure of wound needs to be delayed due to infection, vacuum system can be additionally applied.
|
|
|
|